產品資料

Rapid-Rate Thermal Cycle Chamber

如果您對該產品感興趣的話,可以
產品名稱: Rapid-Rate Thermal Cycle Chamber
產品型號:
產品展商: Lab Companion
產品文檔: 無相關文檔

簡單介紹

Rapid-Rate Thermal Cycle Chamber


Rapid-Rate Thermal Cycle Chamber  的詳細介紹

Rapid-Rate Thermal Cycle Chamber

High temperature rate of change with uniformity and reproducibility

The Rapid-Rate Thermal Cycle Chamber is ideal for test requiring quick changes 15°C/min ramp rate in specimen temperature and 23°C/min ramp rate for air temperature(temperature heat up average). It covers various applications from JEDEC, IEC standards to screening.

It features new technologies, such as a specimen temperature control function maintaining linear specimen temperature change rates during rapid thermal cycling, and temperature ramp control.

Specifications
Model Temperature range Inside / Outside dimensions (mm)
T/270/70/25 -70 to +180°C (-94 to +356°F) W580×H750×D615 / W895×H1950×D2080

Compatible Test standards

IEC-60749-25 : Semiconductor devices – Temperature cycling

IEC-60068-2-14 Nb : Environmental testing – Change of temperature

IEC-61147-5 : Liquid crystal and solid-state display devices – Environmental, endurance and mechanical test methods

JESD22-A105-B : Solid State Devices, testing quality and reliability – Power and temperature cycling

SAE-J1211 : Automotive components - Recommended environmental practices for electronic equipment design

IPC-9701 : Performance test methods and qualification requirements for surface mount solder attachments

Uniformity, reproducibility
Wind simulation allows achieving minimum temperature variations of the specimen, for accurate and quick temperature changes. Specimen temperature ramp rate is 15°C/min, air temperature ramp rate is 23°C/min(temperature heat up average).
Greater temperature ramp control

High precision of the specimen temperature control achieved thanks to a sensor for temperature specimen measurement, and the embedded high-speed controller.

In addition:

  • refrigeration capacities increased at low temperatures,
  • minimization of the difference between test area and specimen temperatures,
  • airflow speed uniformity,

lead to excellent temperature ramp control.

Selection of the temperature control mode

Two temperature control modes are available:

  • A specimen temperature control mode meeting ramp rate conditions of JEDEC standard (15°C/min.)
  • An air temperature control mode for temperature cycle tests.
Free access
Cable ports are equipped on both left and right side, to ease specimen access and wiring.
TEST STANDARD (T/270/70/25, COMPATIBILTY)
Test standard Temperature setting Temperature
change rate
Soak time Number
of cycles
High temperature (°C) Low temperature (°C)
IEC 60749-25 
(JESD22-A104-D)
G +125 (+15, -0) -40 (+0, -10) Specimen temperature,
15°C / min. or less
1, 5, 10, 15 min. Not specified
I +115 (+15, -0) -40 (+0, -10)
J +100 (+15, -0) 0 (+0, -10)
K +125 (+15, -0) 0 (+0, -10)
L +110 (+15, -0) -55 (+0, -10)
N +80 (+15, -0) -30 (+0, -10)
O +125 (+15, -0) -25 (+0, -10)
IEC 60068-2-14 Nb 
(JIS C 60068-2-14 Nb)
+175 ±2
+155 ±2
+125 ±2
+100 ±2
+85 ±2
+70 ±2
+55 ±2
+40 ±2
+30 ±2
-65 ±3
-55 ±3
-40 ±3
-25 ±3
-5 ±3
+5 ±3
1±0.2°C / min.
3±0.6°C / min.
5±1.0°C / min.
(AVG)
Average for up to fi ve minutes
3 hours,
2 hours,
1 hour,
30 min.,
10 min.
3 hours if not specified in product specifi cations
2
IEC-61747-5 
(EIAJ ED-2531A)
+100 ±2
+95 ±2
+90 ±2
+85 ±2
+80 ±2
+75 ±2
+70 ±2
+65 ±2
+60 ±2
+55 ±2
+50 ±2
+45 ±2
+40 ±2
+35 ±2
+30 ±2
-50 ±3
-45 ±3
-40 ±3
-35 ±3
-30 ±3
-25 ±3
-20 ±3
-15 ±3
-10 ±3
-5 ±3
-0 ±3
1±0.2°C / min.
3±0.6°C / min.
5±1.0°C / min.
(AVG)
Average for up to fi ve minutes
3 hours,
2 hours,
1 hour,
30 min.,
10 min.
3 hours if not specified in product specifi cations
2
JESD22-A105-B A +85 (+10, -0) -40 (+0, -10) 6.25°C / min. 10 min. 1000
B +125 (+15, -0) -40 (+0, -10) 5.5°C / min.
IPC-9701 TC1 100 0 Specimen temperature,
20°C / min. or less
Specimen temperature,
10 min.
200
500
1000
3000
6000
TC2 100 -25
TC3 125 -40
TC4 125 -55
TC5 100 -55
IPC-TM-650 2.6.6 A +125 (+3, -0) -65 (+0, -5) - 30 min. 5
B +85 (+3, -0) -55 (+0, -5)
SAE-J1211 +85 to +150 -40 4 to 6°C / min. Low temperature,
4 hours
-